Voltage measurement |
4.0 ... 600.0 V RMS value ± 0.5% of the measured value ± 5 digits / 0.1 V |
Voltage peaks (> 10V) |
50 Hz - 60 Hz / ± 5% ± 50 digits |
Current measurements |
4.0 ... 1500.0 A real ± 1% V. of the measured value ± 5 digits / 0.01 A |
Current peaks (> 20A) |
50 Hz - 60 Hz / ± 5% ± 50 digits |
Active power P |
10 W ... 9999 kW in 5 measurement ranges
± 1% of the value + 20 digits (>20 V y >20 A)
± 2% of the value + 40 digits (>20 V y >20 A)
0.1 W ... 1 kW according to the measurement range |
Apparent power S |
10 VA ... 9999 kVA in 5 measurement ranges
± 1% of the value + 20 digits (>20 V y >20 A)
± 2% of the value + 40 digits (>20 V y >20 A)
0.1 VA ... 1 kVA according to the measurement range |
Reactive power Q |
10 VAr ... 9999 kVAr in 5 measurements ranges
± 1% of the value + 20 digits (>20 V y >20 A)
± 2% of the value + 40 digits (>20 V y >20 A)
0.1 VAr ... 1 kVAr according to the measurement range |
PF power factor |
0.000 ... 1.000
± 0.04 (>20 V >20 A) / 0,001
± 0.1 (>20 V >20 A) / 0,001 |
Phase angle |
-180 ... +180 / ±1 / 0,1
0 ... +360 / ±1 / 0,1 |
Frequency measurement (U > 50 V) |
46 ... 65 Hz
±0.3 Hz / 0.1 Hz |
Active energy |
0 ... 999.999 kWh
± 1% of the value + 20 digits |
Apparent energy |
0 ... 999.999 kVAh
± 1% of the value + 20 digits |
Reactive energy |
0 ... 999.999 kVarh
± 1% of the value + 20 digits |
Harmonics in the voltage (relative)
50 - 60 Hz > 80 V AC |
1 - 20 th / ± 2% / 0.1%
21 - 50 th / ± 4% of the value ± 2% / 0.1% |
Harmonics in the voltage (absolute)
50 - 60 Hz > 80 V AC |
1 - 20 th / ± 2% / 0.1%
21 - 50 th / ± 4% of the value ± 0.5% / 0.1% |
Harmonics in the current (relative)
50 - 60 Hz > 80 V AC |
1 - 20 th / ± 2% / 0.1 %
21 - 50 th / ± 4% of the value ± 2% / 0.1 % |
Harmonics in the current (absolute)
50 - 60 Hz > 80 V AC |
1 - 20 th / ± 2% of the value ± 0.4 A / 0.1 A
21 - 50 th / ± 4% of the value ± 0.4 A / 0.1 A |
Total harmonic distortion (THD-F)
50 - 60 Hz > 80 V y > 20 A |
0.0 - 20% / 2% / 0.1%
20.1 - 100% / ± 6% of the value ± 1% / 0.1%
100.1 - 999.9 % / ± 10% of the value ± 1% / 0.1% |
Max. clamp opening |
55 mm / 2.2 in |
Range selection |
Auto |
Overload indication |
"OL = Overload |
Memory |
50,000 readings |
Interface |
USB |
Software and data cable |
included in the delivery. They can be used with
Win 2000, XP, ME |
Display |
graphic LCD display, 128 x 64 with backlight |
Power supply |
2x 1.5V AA batteries |
Energy consumption |
Approx. 10-mA |
Dimensions |
271 x 112 x 46 mm / 10.6 x 4.4 x 1.8 in |
Weight |
650 g / 1.4 lbs with batteries |
Operating conditions |
85% max. R.H. / -10 ... 50°C / 14 ... 122°F |
Protection / Standards |
CAT III 600 V / EN 61010-2-032 |