Size Range |
0.5 to 55.0 µm |
Size Channels |
Factory calibrated at 0.5, 1.0, 2.5, 5.0, 10.0, 20.0 µm variable binning |
Flow rate |
0.1 CFM (2.83 LPM) |
Concentration Limit |
10,000,000 Particles/ft³ @ 10% coincidence loss |
Battery Run Time |
10 hours |
Light Source |
Long life laser diode |
Counting Efficiency |
50% @ 0.5 µm; 100% for particles >0.75 µm per JIS |
Zero Count |
<1 count / 5 minutes (<2 particles/ft³) (per ISO 21501-4 & JIS) |
Count Modes |
Automatic, manual, real-time meter, cumulative/differential, mass concentration, count or concentration |
Count Alarms |
1 to 9,999,999 counts |
Calibration |
NIST traceable |
Display |
4.3" (10.9 cm) WQVGA (480x272) color touch screen |
Printer (Optional) |
External thermal printer. |
Vacuum Source |
Internal pump with automatic flow control |
Filtered Exhaust |
Internal HEPA filter |
Number of Channels |
6 |
Custom Size Channels |
Calibration for custom size channels available |
Audible Alarm |
Adjustable built-in alarm |
Battery |
Removable Li-ion |
Battery Recharge Time |
4 hours within instrument (<2 hours with external battery charger) |
Reports |
ISO 14644-1, EU GMP Annex 1, FS 209E |
Recipes |
50 user-configurable recipes |
Communication Modes |
Ethernet and USB |
Optional Communication Mode |
Wireless 802.11 b/g, RS485 or RS232 |
Environmental Sensor |
Includes temperature and relative humidity probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC),
15-90% ±2% relative humidity |
Alarm |
Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow |
Standards |
ISO 21501-4 and JIS B9921 |
Calibration |
Recommended minimum once per year |
External Surface |
High impact injection molded plastic |
Dimensions (L x W x H) |
10” x 5” x 4.5” (25.4 cm x 12.9 cm x 11.4 cm) includes handle and does not include probes |
Weight |
2.2 lb (1.0 kg) |
Accessories |
Operating manual on USB flash drive, isokinetic probe, temperature relative humidity sensor, purge filter, battery, data download software, USB cable, power supply & cable |
Optional Accessories |
Printed manual, carrying case, spare battery, external battery charger, external printer and isokinetic probes |
Buffer Memory |
45,000 sample records (rotating buffer) including particle count data, environmental data, locations and times. Scrollable on screen or printout |
Sample Locations |
Up to 1,000 locations 20 characters long |
Sample Time |
1 second to 99 hours |
Power |
110 to 240 VAC 50/60 Hz universal in-line power supply |
Operating Conditions |
41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing |
Storage Conditions |
32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing |
Warranty |
2 Years. Extended warranties available. |