Liên hệ
The zero and span drift specifications assume that the analyzer is equilibrated, and is at constant temperature, with a properly maintained sensor.

Accuracy

± 2.0% of reading, ± 1 least significant digit. (Limited to accuracy of calibration standard)

Repeatability

± 0.5% of full scale

Linearity

± 1.0% of full scale

Zero Drift

± 1.0% of full scale (24 hours)

Span Drift

Less than ± 2.0% of full scale (24 hours)

Calibration

Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service.

Lag Time

Less than 1 second

Minimum Detectability

1.0% of full scale

Range

0-2000 ppm

Resolution

1 ppm

Gas

Hydrogen (H2)

Number of Detection Points

1

Application

Continuous

Note

 
Specifications