Input |
Number of points |
3, 6, 9, 12 points (select at a time of purchase) *1 |
Measuring cycle |
100 ms |
Recording cycle |
1 sec. to 60 min. (Main/Sub record)
100 ms can be selected only at the Sub record |
Input signal |
Thermocouple: K, E, J, T, C, N, PLII, U, L, R, S, B, PR20-40, Au-Fe
RTD: Pt 100 Ω, JPt 100 Ω
DC voltage: -10 to 10 mV, 0 to 20 mV, 0 to 50 mV, -0.2 to 0.2 V, -1 to 1 V, -10 to 10 V, 0 to 5V
DC current: 4 t0 20 mA (External shunt resistance) |
Performance |
Indication accuracy |
± (0.1 % + 1 digit) of input range |
Display |
LCD |
5.7" TFT(QVGA) with Touch Panel and backlight |
Contents |
- Trend display (in vertical or horizontal) selected in the refreshment cycles of 1 sec. to 60 min.
- Bar graph (refresh cycle: 1 sec.)
- Digital display (refresh cycle: 1 sec.)
- Event summary display (alarm and message summary)
- Ethernet log display
- Historical trend display |
Recording function |
Recording medium |
SD card (max. 2 GB)
SDHC card (max. 32 GB) |
Internal memory |
about 100 MB |
Recording method |
- Writing starts in fixed cycles by turning ON the REC key on the front panel.
- Data is recorded in a new file every time the recording starts. |
Data save cycle |
Selectable with 1 hour to 1 year |
Data format |
CSV 120 bytes per sampling (at 6 channels input)
Binary (Data cannot be read directly into Excel, etc.) 32 bytes per 1 sampling (at 6 channels input) |
Alarm function |
Number of settings |
Up to 4 alarms can be set for each channel. |
Communication |
Ethernet 10 BASE-T |
HTTP server, FTP server and Modbus/TCP protocol |
General |
Mounting method |
Panel-mount (vertical panel) |
Material |
Polycarbonate Glass 10 % UL94-V0 (case, bezel) |
Dimensions |
Panel mount: 150 (W) × 144 (H) ×181.8 (D)mm |
Weight |
about 0.9 kg (3 channels), about 6 kg (6 channels) |
Power requirement |
100V to 240V AC, 50/60 Hz |
Power consumption |
15 VA max. at 100V AC, 25 VA max. at 240V AC) |
Option |
Relay card
DI/DO card |
Relay: 6 points (1a contacts)
DI/DO: input 9 points / output 12 points (open collector) |
Communication |
RS-485 (MODBUS RTU protocol) |