Hotline 24/7: 0986951455
Việt Nam English
Sản phầm nổi bật

Máy dò khí Interscan 4183UL, 4186UL, 4187UL, 4189UL UL classified intrinsically safe Portable Analyzer (4000 Series) - Hydrazine

Mã sản phẩm: 4183UL, 4186UL, 4187UL, 4189UL
Tình trạng:

Máy dò khí Interscan 4183UL, 4186UL, 4187UL, 4189UL UL classified intrinsically safe Portable Analyzer (4000 Series) - Hydrazine

Liên hệ
Liên hệ

4183UL

 UL Classifed Intrinsically Safe for Class I, Groups A, B, C and D

Accuracy

± 15% of full scale (Limited to accuracy of calibration standard)

Repeatability

± 10% of full scale

Minimum Detectability

10% of full scale

Linearity

± 1.0% of full scale

Zero Drift

± 10% of full scale (24 hours)

Span Drift

Less than ± 10% of full scale (24 hours)

Calibration

Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service.

Lag Time

Less than 1 second

Range

0-0.5/0-2 ppm

Resolution

1% of full scale

Dimensions

7" H x 4" W x 8.875" D   (178 x 102 x 225 mm)

Weight

4.5 lb    (2.0 kg)

Gas

Hydrazine (N2H4)

Application

Portable

Note

The zero and span drift specifications assume that the analyzer is equilibrated, and is at constant temperature, with a properly maintained sensor.

Special Note

Certain specification items on this model reflect performance different from most other Interscan analyzers, but are in accordance with NASA guidelines, and the extreme difficulties encountered in low-level hydrazine measurement.

Intrinsic safety

 
Specifications  

4186UL


UL Classifed Intrinsically Safe for Class I, Groups A, B, C and D

Accuracy

± 2.0% of full scale (Limited to accuracy of calibration standard)

Repeatability

± 0.5% of full scale

Minimum Detectability

1.0% of full scale

Linearity

± 1.0% of full scale

Zero Drift

± 1.0% of full scale (24 hours)

Span Drift

Less than ± 2.0% of full scale (24 hours)

Calibration

Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service.

Lag Time

Less than 1 second

Range

0-2/0-10 ppm

Resolution

1% of full scale

Dimensions

7" H x 4" W x 8.875" D   (178 x 102 x 225 mm)

Weight

4.5 lb    (2.0 kg)

Gas

Hydrazine (N2H4)

Application

Portable

Note

The zero and span drift specifications assume that the analyzer is equilibrated, and is at constant temperature, with a properly maintained sensor.

Special Note

-----

Intrinsic safety

 
Specifications  

4187UL


UL Classifed Intrinsically Safe for Class I, Groups A, B, C and D

Accuracy

± 15% of full scale (Limited to accuracy of calibration standard)

Repeatability

± 10% of full scale

Minimum Detectability

10% of full scale

Linearity

± 1.0% of full scale

Zero Drift

± 10% of full scale (4 hours)

Span Drift

Less than ± 10% of full scale (4 hours)

Calibration

Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service.

Lag Time

Less than 1 second

Range

0-100/0-2000 ppb

Resolution

1% of full scale

Dimensions

7" H x 4" W x 8.875" D   (178 x 102 x 225 mm)

Weight

4.5 lb    (2.0 kg)

Gas

Hydrazine (N2H4)

Application

Portable

Note

The zero and span drift specifications assume that the analyzer is equilibrated, and is at constant temperature, with a properly maintained sensor.

Special Note

Certain specification items on this model reflect performance different from most other Interscan analyzers, but are in accordance with NASA guidelines, and the extreme difficulties encountered in low-level hydrazine measurement.

Intrinsic safety

 
Specifications  

4189UL

 UL Classifed Intrinsically Safe for Class I, Groups A, B, C and D

Accuracy

± 15% of full scale (Limited to accuracy of calibration standard)

Repeatability

± 10% of full scale

Minimum Detectability

10% of full scale

Linearity

± 1.0% of full scale

Zero Drift

± 10% of full scale (4 hours)

Span Drift

Less than ± 10% of full scale (4 hours)

Calibration

Against standard gas mixture, or via INTERSCAN's Electronic Calibration Service.

Lag Time

Less than 1 second

Range

0-100/0-1000 ppb

Resolution

1% of full scale

Dimensions

7" H x 4" W x 8.875" D   (178 x 102 x 225 mm)

Weight

4.5 lb    (2.0 kg)

Gas

Hydrazine (N2H4)

Application

Portable

Note

The zero and span drift specifications assume that the analyzer is equilibrated, and is at constant temperature, with a properly maintained sensor.

Special Note

Certain specification items on this model reflect performance different from most other Interscan analyzers, but are in accordance with NASA guidelines, and the extreme difficulties encountered in low-level hydrazine measurement.